Inviro EZ Wing Needless IV Access Instructions for Use

IV Injection Site Access

Medication Preparation: To access split septum or needlefree connector IV injection sites use the InviroLINK>> in combination with the Inviro Luer Lock Syringe with InviroSTRIPE (Inviro LL Syringe) or standard Luer Lock syringe:

  1. Peel package as indicated to allow proper removal of InviroLINK>>. (See Figure 1).
  2. Attach InviroLINK>> onto the Inviro LL Syringe or standard Luer Lock syringe.
  3. Prepare vial for medication draw up.
  4. Penetrate center of vial with InviroLINK>>. (See Figure 2).
  5. Draw up medication per facility protocol.

Split Septum IV Injection Site:

  1. Insert InviroLINK>> attached to Inviro LL Syringe or standard Luer Lock Syringe into center of split septum IV injection site for medication injection. (See Figure 3).
  2. After completion of medication delivery, discard immediately in sharps container per facility protocol.

Needlefree Connector IV Injection Site:

  1. Remove InviroLINK>> and dispose of per facility protocol.
  2. Connect Inviro LL Syringe or standard Luer Lock syringe into needlefree connector IV injection site.

Figure 1Learn more about needleless IV access.

Figure 2Figure 2

Figure 3Figure 3

Caution

This device is sterile unless the package has been damaged or opened.
If unable to activate safety mechanism, discard syringe immediately into an approved
sharps container.
This device is a single use item only.
InviroLINK>> is designed for penetration into vials and split septum IV injection sites
only, not for patient injection.
Maximum needle length not to exceed 1 1/2" or 38.1mm.

Note: InviroSTRIPE, the integral write-on Stripe, allows you to annotate information directly onto the syringe barrel with an ink cure time of 30 seconds.

SNAP!® syringe patents US: 5,263,933, 5,462,531, 5,520,649, 6,033,386, 6,117,113, 6,344,031, 6,616,632 and additional foreign patents and patents pending. Additional applicable patents: US: 5263934, 5466226, 5380286, 5364359 and 6315760; Canada: 131718. Caution; US Law restricts this device to sale by or on the order of a physician.